Quantum Scalar i40 Guida Utente Pagina 319

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 318
Appendix B: Tape Alert Flag Descriptions
Quantum Scalar i40 and Scalar i80 User’s Guide 291
33 Eject media C The operation has failed:
1 Eject the tape or magazine.
2 Insert the tape or magazine again.
3 Restart the operation.
Error recovery action.
34 Microcode
update fail
W The microcode update has failed
because you have tried to use the
incorrect microcode for this tape drive.
Obtain the correct microcode and try
again.
Microcode update
failed.
35 Drive humidity W Environmental conditions inside the
tape drive are outside the specified
humidity range.
Drive humidity limits
exceeded.
36 Drive
temperature
W Environmental conditions inside the
tape drive are outside the specified
temperature range.
Cooling problem.
37 Drive voltage W The voltage supply to the tape drive is
outside the specified range.
Drive voltage limits
exceeded.
38 Predictive
failure
C A hardware failure of the tape drive is
predicted. Call the tape drive supplier
help line.
Predictive failure of
drive hardware.
39 Diagnostics
required
W The tape drive may have a hardware
fault. Run extended diagnostics to
verify and diagnose the problem. Check
the tape drive user’s manual for device-
specific instructions on running
extended diagnostic tests.
The drive may have a
hardware fault that
may be identified by
extended diagnostics
(i.e.,
SEND
DIAGNOSTIC
command).
40
46
Obsolete
47
49
Reserved
No. Flag Severity
Recommended Application Client
Message Probable Cause
Vedere la pagina 318